Specifications/Details :
- Large LCD display waveform and data
- High-resolution thermal printing array system
- Frequency response is as high as 150Hz
- Capable of printing continuously on channel trace and annotations including lead mark and parameter such as sensitivity, paper speed and filter operation status
- High resolution digital filter inhibits baseline drift without affecting ECG waveform
- Concept of floating input circuit meets IEC safety standard
- Rechargeable battery supporting more than 50 patient examinations
| Specification: | |
|---|---|
| Input Circuit | Floating; Protection against Defibrillator effect |
| Lead | Standard 12 Leads, Lead change automatically |
| Patient Current Leakage: | ≤10μA |
| Calibrating Voltage: | 1mV±3% |
| Frequency Response: | 0.05Hz~150Hz (IEC) CMRR>100dB |
| Paper Speed: | 25, 50mm/s±3% |
| Recording Paper: | 50mm,20m, high-speed roll |
| Input Circuit Current: | ≤50mA |
| Sensitivity: | 1/2, 1, 2 (cm/mV) |
| Skin Voltage Tolerance: | ±300mV |
| Dimensions: | 310mm×230mm×70mm |
| Input Impedance: | ≥50ΜΩ |
| A/D Conversion: | 12 bit |
| Time Constant: | >3.2seconds |
| EMG Filter: | 35Hz (-3dB) / 25Hz (-3dB |
| Safety Standard: | IEC class I, type CF |
| Noise Level : | <15μVp-p |
| Weight | 2.75kg |
Standard Accessories:
- Patient Cable,Thermal Printing Paper, Limb Electrodes, Chest Electrodes, Power Supply Cable, Grounding Wire